Design of a novel DXA scanner with a CdTe photon-counting Timepix4 detector for peripheral bone densitometry

L.A. Cerbone, J. Žemlička, B. Bergmann, P. Smolyanskiy, P. Mánek, G. Mettivier, L. Cimmino, Y. Lai, X. Jia, S.K. Boyd, P. Russo

Abstrakt

Bone densitometry in osteoporosis diagnosis via dual-energy X-ray absorptiometry (DXA) can benefit from advances in imaging detector technology. We devised a com-pact imaging scanner – DXA4A – using a photon counting and energy-sensitive Timepix4 hybrid pixel detector (512 x 448 pixels, 55 µm pitch), for areal bone mineral density (aBMD) assessments in the distal radius and tibia in the clinic and for future in-flight astronauts’ bone health assessment. We present the design and initial tests of the scanner. A Timepix4 detector with a 1-mm-thick CdTe sensor was tested in the la-boratory with X-ray tube sources, acquiring first images of test samples. Monte Carlo simulations were implemented for scanner design and performance prediction, using 50 kVp unfiltered and 100 kVp Sm K-edge filtered spectra. With a digital twin of the scanner and patient wrist we set-up a virtual imaging study and determined the aBMD in the forearm of a patient (0.515 ± 0.048 g/cm2), in agreement with the clinical DXA value (0.571 g/cm2 for the total forearm). This study highlights the feasibility of realiz-ing a compact DXA scanner for distal tibia and radius with spectral capabilities, ex-ploiting Timepix4 hybrid detectors for its peculiar energy sensitivity and photon event timing properties for tissue identification.