Improved algorithms for determination of particle directions in space with Timepix3
Abstrakt
Timepix3 pixel detectors have demonstrated great potential for tracking applications. With 256 × 256 pixels, 55 µm pitch and improved resolution in time (1.56 ns) and energy (2 keV at 60 keV), they have become powerful instruments for characterization of unknown radiation fields. A crucial pre-processing step for such analysis is the determination of particle trajectories in 3D space from individual tracks. This study presents a comprehensive comparison of regression methods that tackle this task under the assumption of track linearity. The proposed methods were first evaluated on a simulation and assessed by their accuracy and computational time. Selected methods were then validated with a real-world dataset, which was measured in a well-known radiation field. Finally, the presented methods were applied to experimental data from the Large Hadron Collider. The best-performing methods achieved a mean absolute error of 1.99° and 3.90° in incidence angle θ and azimuth φ, respectively. The fastest presented method required a mean computational time of 0.02 ps per track. For all experimental applications, we present angular maps and stopping power spectra.
Klíčová slova
- Analysis and statistical methods
- Data processing methods
- Data reduction methods
- Pattern recognition
- Cluster finding
- Calibration and fitting methods
- Particle tracking
- Pixel detectors
- Timepix3
- track reconstruction
- regression methods
- pixel detector
- radiation field characterization
- Large Hadron Collider
- angle of incidence
- azimuth angle