Detector response and performance of a 500 mu m thick GaAs attached to Timepix3 in relativistic particle beams

B. Bergmann, P. Azzarello, P. Broulim, P. Burian, L. Meduna, S. Pospíšil, L. Tlustos

Abstrakt

The performance and response of a Timepix3 detector with a 500 mu m thick GaAs:Cr sensor layer was investigated in different radiation fields. The sensor resistivity was rho approximate to 10(9) Omega cm. Fitting different modified Hecht functions, which take the small pixel effect into account, the mobility-lifetime products of mu(e)tau(e) = (0.773 +/- 0.018 degrees) x 10-4 cm(2)V(-1) and mu(e)tau(e) = (0.996 +/- 0.056) x 10(-4) cm(2)V(-1) were determined. Hereby, the latter value is favored due to the better agreement of fit and data. In a measurement in a 40 GeV/c pion beam, the drift times and charge collection efficiencies were studied as a function of the interaction depth. We present the measured drift velocity as a function of the electric field strength and compare the determined dependence with a model. In a measurement in a mixed ion beam, we study the capability of the detector to separate different ion species. We show the detector response in the form of tracks and discuss heavy ion track features.

Klíčová slova

Konference

21st International Workshop on Radiation Imaging Detectors 2019-07-07