Environmental radionuclides as contaminants of HPGe gamma-ray spectrometers: Monte Carlo simulations for Modane underground laboratory
Abstrakt
The main limitation in the high-sensitive HPGe gamma-ray spectrometry has been the detector background, even for detectors placed deep underground. Environmental radionuclides such as K-40 and decay products in the (238)u and Th-232 chains have been identified as the most important radioactive contaminants of construction parts of HPGe gamma-ray spectrometers. Monte Carlo simulations have shown that the massive inner and outer lead shields have been the main contributors to the HPGe-detector background, followed by aluminum cryostat, copper cold finger, detector holder and the lead ring with FET. The Monte Carlo simulated cosmic-ray background gamma-ray spectrum has been by about three orders of magnitude lower than the experimental spectrum measured in the Modane underground laboratory (4800 m w.e.), underlying the importance of using radiopure materials for the construction of ultra-low-level HPGe gamma-ray spectrometers.
Klíčová slova
- ACCELERATOR MASS-SPECTROMETRY
- BACKGROUND CHARACTERISTICS
- LEVEL
- TECHNOLOGIES
- DETECTOR
- RADIOACTIVITY
- SPECTROSCOPY
- RADIOMETRICS
- SYNERGY
- AMS
- environmental radionuclides
- radioactive contamination
- detector background
- cosmic ray background
- lead shielding
- potassium-40
- cryostat
- underground laboratory
- Modane underground laboratory