Particle telescope with Timepix3 pixel detectors

P. Burian, P. Broulim, V. Georgiev, B. Bergmann, S. Pospíšil

Abstrakt

In the presented work, we demonstrate how Timepix3 detectors, featuring the Katherine (Gigabit Ethernet interface) readout, can be used in a particle telescope. For the highly precise timestamp synchronization, a special timing unit (with time resolution similar to 13 ps) based on a Time-to-Digital-Converter (TDC) is used and interconnected with the readouts. The designed concept of the measurement chain was tested at the SPS facilities at CERN in a 40 GeV/c pion beam using a telescope of 3 synchronized Timepix3 detectors. Evaluating spatially and temporally coincident events, an overall time resolution below 2 ns was determined, whereby a systematic error of < 1 ns was present.

Klíčová slova

Konference

19th International Workshop on Radiation Imaging Detectors Krakow, Poland · 2017-07-02