Study of the Capabilities of the Timepix Detector for Ion Beam Radiotherapy Applications
Abstrakt
The finite range of ion beams in matter and the
presence of the Bragg-peak at the end of their range enable
to create highly localized dose distributions. This advantage
represents at the same time a challenge - namely to control
all factors influencing the dose distribution with high precision.
To ensure the required quality of the dose distribution applied to
the patient, a number of measurement procedures is conducted.
Many of these techniques were adopted from photon beam
radiotherapy. Improvements in terms of the information amount
gained are expected from new generations of dedicated ion detection
techniques. One of the possible research directions is given
by semiconductor-based detectors, which are at present rarely
used in ion beam radiotherapy. In this contribution we report on
the studies of the capabilities of the Timepix detector. This hybrid
semiconductor pixelated detector was developed by the Medipix
Collaboration at CERN. Its high spatial resolution allows an
online registration of single ions. Experiments were performed
at the Heidelberg Ion Beam Therapy Center in Germany using
carbon ion beams. The signal pattern of single ions in the detector
signal was shown to provide separation of the primary and
different types of secondary ions. The direction of the secondary
ions emerging from different irradiated phantoms were measured
using 3D voxel detector containing several Timepix layers. The
analysis of the recorded ion tracks shows a correlation with the
beam range, position and width, which are all of high interest
for monitoring of the beam within the patient during the therapy
delivery. The results of the discussed studies show that the
Timepix detector offers attractive capabilities for the needs of
carbon ion beam therapy.