Towards Fragment Distinction in Therapeutic Carbon Ion Beams: A Novel Experimental Approach Using the Timepix Detector

B. Hartmann, K. Gwosch, C. Granja, J. Jakůbek, S. Pospíšil, O. Jaekel, M. Martisikova

Abstrakt

Radiotherapy with carbon ion beams is a highly
precise method for cancer treatment. This is due to the finite
range and the relatively low lateral scattering of the ions.
However, on their path through tissue, the carbon ions may
undergo nuclear fragmentation, resulting in lighter but energetic
projectile fragments. Since the biological effect of the fragments
differs from the primary particles, it is important to consider
fragmentation in beam models used for therapy planning.
Until now, large apparatus have been utilized for ion spectroscopic
measurements. Employing a small detector would allow
investigations directly within therapy relevant phantoms.We used
the Timepix detector, consisting of a 300m thick silicon sensor
bump-bonded to a readout chip with 256 x 256 square pixels
(55m pitch). The high spatial resolution enables to visualize
and distinguish tracks of single particles. The pixel-wise energy
calibration allows to directly determine the particle energy loss
in the sensitive layer. Charge released in the detector by an ion
spreads out during charge collection and is collected by several
adjacent pixels forming signal clusters.
Experiments were performed at the Heidelberg Ion-Beam Therapy
Center, using a carbon ion pencil beam of E=271 MeV/u. To
investigate fragments in different material depths, the detector
was placed perpendicular to the beam with 12 to 45 cm thick
PMMA slabs in front of it. Pattern recognition analysis of
the signal showed a clear dependence of cluster parameter
distributions on the PMMA depth. This was used to identify
the particular particle species in the obtained spectra. In this
way, discrimination between primary carbon ions and hydrogen,
helium and heavier fragments is possible.
The presented novel method enables fragment distinction in
mixed particle fields. Its main advantage lies in the flexibility
of the set-up.