Metal micro-detectors for radiation therapy instrumentation

O. Kovalchuk, A. Pugatch, A. Chaus, O. Fedorovich, O. Okhrimenko, D. Storozhyk, M. Campbell, L. Tlustos, X. Llopart, S. Pospíšil, Y. Prezado, M. Renier

Abstrakt

Physics and techniques of the metal detectors measuring and imaging charged particles and synchrotron radiation beams are presented. An extremely low thickness (~ 1 micron) of the Metal Micro-strip Detector (MMD) combined with its high radiation tolerance (~ 100 MGy) introduces an opportunity to keep a device in the beam, permanently. 'In-situ' operation of the MMD provides non-destructive beam diagnostics in real time. Characterization studies of the MMD at the Mini-beam Radiation Therapy setup (Bio-Medical Beamline ID17, ESRF, France) have been recently performed. The results obtained illustrate an excellent performance of the metal micro-pixel detector TimePix providing online synchrotron radiation beam monitoring.

Klíčová slova

Konference

IEEE Nuclear Science Symposium and Medical Imaging Conference 2011 Valencia · 23.10.2011