Monitoring of Ion Beam Energy by Tracking of Secondary Ions: First Measurements in a Patient-like Phantom
Abstrakt
Due to the finite range of ions in matter and the
presence of the Bragg-peak, ion beams provide highly localized
dose distributions. In radiation therapy with ion beams, unpredictable
changes within the patient can deteriorate the quality
of the dose distribution in the target. Therefore it is desired
to monitor the beam within the patient in a non-invasive way.
In this contribution the information carried by secondary ions,
which are emerging from a patient during carbon ion beam
irradiation, is studied. In a framework of investigations on the
feasibility of beam monitoring exploiting secondary ions, the
aim of this study was to show the possibility of distinction of
two different primary carbon ion beam energies solely from the
analysis of the secondary ion directions measured. The situation
was complicated by realistic tissue inhomogeneities simulated by
using a patient-like phantom. Experiments were performed at
the Heidelberg Ion Beam Therapy Center in Germany using
narrow carbon ion beams. The emerging secondary ions were
registered by the Timepix detector. To determine the direction
of the particles, a multi-layered detector (3D voxel detector)
was employed. Clear differences between the application of the
two beams with different energies (213 and 250 MeV/u) could
be observed in the distributions of the measured secondary
ions, despite the inherent tissue inhomogeneities. This result was
achieved in both brain (homogeneous) and skull base regions
(containing inhomogeneities). Differences between the energies
could be observed with the detector placed on different positions
around the head. The performed first experiments towards 12C
beam monitoring in a patient-like geometry exploiting tracking
of secondary ions with a 3D voxel detector show, that the
information about the therapeutic carbon ion beam which is
carried by secondary ions is promising and worthwhile to be
investigated further.
Klíčová slova
Konference