3D Beam Monitoring for 12C Radiotherapy by Tracking of Secondary Ions Using the Timepix Detector
Abstrakt
Radiotherapy with narrow 12C ion beams enables
treatment of tumors with high precision while sparing the surrounding
healthy tissue. Unpredictable changes in the patient's
geometry can alter the ion range and result in changes in the
dose distribution. Therefore, it is desirable to verify the actual
dose delivery in the patient, preferably in real-time and in a
non-invasive manner.
Currently, the only technically feasible dose delivery monitoring
method is based on tissue activation measurements by
means of positron emission tomography (PET). As an alternative
to PET-based measurements, beam range monitoring exploiting
the detection of prompt secondary ions has been suggested. This
modality is expected to allow for real-time monitoring, thereby
reducing the influences of physiological signal wash-out known
from PET-based techniques.
In this contribution, the potential of monitoring narrow 12C ion
beams in a head-sized PMMA phantom by tracking of secondary
ions is investigated experimentally. Experiments with therapeutic
carbon ion beams were performed at the Heidelberg Ion Beam
Therapy Center (HIT), Germany. The Timepix detector was used
to track secondary ions emerging from the irradiated phantom.
Analysis of the secondary ion directions was used to monitor the
range, width and position of pencil-like 12C ion beams in therapy
relevant conditions.
Clear dependences of the secondary ion track distribution on
the investigated beam settings were found. Detectable were beam
range differences down to about 2mm and differences in the
beam width of 1.4 mm. Furthermore, lateral shifts of the beam
position by 1mm were measurable. The presented experiments
show the potential of secondary ion tracking for monitoring
therapeutic carbon ion beams.