The proposed project concerns the research and development of highly efficient neutron
detectors with spatial resolution on the level of 50-100 µm for neutron cameras and portable dosimeters usable in mixed radiation field by using silicon strip sensors with a pitch of about 75 µm and an area up to 5x5 cm2. For that purpose the single strip sensors will be connected to the corresponding R/O electronics and characterized by means of alpha, beta and gamma radiation, and by neutrons in case of sensors covered by a neutron converter. To verify the fabrication of the proposed sensors and the optimisation of complex assembly and packaging of the detection system using state-of-the-art silicon and micro-machining technology. To demonstrate a functionality of a new detector design concepts (stacking and 3-dimensional micro-structures), the stack of two sensor layers interlaced with 6LiF converter will be measured and results will be compared to existing technology. To verify the potential of the proposed sensors to measure mixed radiation field with improved spatial resolution. To perform experimental in-depth characterisation of the proposed sensors through state-of-the-art assembly and custom readout electronics.

Původní abstrakt

The main aim of the project is to develop a portable highly efficient and large area neutron
sensitive detecting device for measurement of composition and spectral characteristics of
mixed radiation fields. The device development tends to use in environmental and health
research with a special attention to radiation protection and medical use of radiation. It is
also proposed to operate a system of multiple detectors in network controlled via a web
interface. The project is based on established previous cooperation of all partners that are
involved with already well-defined roles. For the project purpose a new compact multilayer
silicon strip detector based on technologies and know-how on strip and pixel detectors
available at SINTEF, and on experience with a broad exploitation of these detectors in IEAP
CTU will be designed and developed. To read the signals from the compact multilayer strip
detectors, the ASIC electronic chips will be used in connection to appropriate interfaces
controlled by software developed in cooperation of IDEAS and IEAP CTU. Dosimetric partners
from NRPI and CTU represent final users of the developed device. They will take care about
device characterization, testing and calibration for specific applications within different
radiation conditions.